Input Queue General Information

New Input Queue (Cut-in on serial # P1264 and above)

The new Input Queue has the following:

Description

This module accepts new MTUs from the user and prepares them for handoff to the Distributor.
The module has the capacity to hold 25 MTUs. In addition, the module holds and controls the MTU barcode scanner/

Interfacing Modules

A single 6-pin cable connects the MTU Input Queue module to the COP.
The module interfaces with the Distributor by packing the MTUs and placing them, one-at-a-time, in a position for the Distributor to grab. In addition, the module holds and controls the MTU barcode scanner, which is triggered twice for each MTU—once immediately after picking a new MTU and once immediately before going to the Luminometer.

Theory/Sequence of Operation

With the MTU Input Queue drawer open, an operator places a box of MTUs and loads 10 MTUs onto the rails.
If necessary, the operator may load additional MTUs until full.
When the drawer is closed, the MTU pusher moves forward, packing the MTUs together until the final MTU depresses the packing sensor button.
Using the encoder on the packing motor, the module is able to determine how many MTUs are in the module.